FOR USE WITH:
- Chromium Connect, PN-1000171
 - Chromium Connect, PN-1000180
 
v2025.03.31.01 (April, 2025)
Features/Enhancements:
- Improved instrument status reporting in telemetry data
 
Bug Fixes & Changes:
- Resolved gantry mount tips failure that impacted aspiration
 - Resolved login failure to ensure HEPA functionality (if available) during a run
 - Resolved troubleshooting package error, ODTC logs are now included
 
v2024.09.30.01 (October, 2024)
Features/Enhancements
- Added functionality to the Troubleshooting screen for collecting Thermal Cycler logs
 - Improved plate movement checks for failure detection
 - Now displaying the deck and plate location of samples when an error or abort occurs for easier troubleshooting
 - Added a new button to refresh the wireless network list
 - Added a new Assay Setup screen to enable the selection of subsequent assay(s) to run after the current assay
 
Bug Fixes & Changes
- Resolved issue where the Network Connection Screen Wi-Fi SSID dropdown did not display all available networks.
 - Improved Liquid Handler error recovery and reporting
 - Added the option to display password in plain text
 
v2024.03.29.01 (April, 2024)
Features/Enhancements
- Changed UI references to “Maintenance” with new “Cleaning” and “System Checks”
 - New “System Check” Notification when starting an Assay
 - New “Cleaning” Notification at login
 - New customer setting to change the PM notification periodicity
 
Bug Fixes & Changes
- Discontinued Single Cell 3’ Single Index Plate usage
 - Fixed External Network Connection’s inability to switch to and from DHCP
 - Fixed External Network Connection’s inability to enter a proper static IP Address
 
v2023.09.30.05 (October, 2023)
Features/Enhancements
- New Telemetry Server
 - New service user software tools screen
 - Functional test that utilizes new Chromium Automated Test Library and Gel Bead Kit
 - New user setting for instruction level
 - New service user Layout Manager Tool
 - New notification during login when Nimbus device is powered OFF
 
Bug Fixes & Changes
- CPAC driver termination time reduced
 - V(D)J Amp Kit final loading screen now displays the correct primer tube image
 - CORE Paddle and CORE Tip pickup failure no longer creates infinite retries
 
v2023.03.31.01 (June, 2023)
Features/Enhancements
- New Fixed RNA Library Construction Assay
 - New assay workflow to process “Purified DNA Feature Barcode” to a Feature Barcode Library
 - New Single Cell 3’ and Single Cell 5’ high volume sample input (32 μl) capability
 - Improved sample index plate labeling
 - Improved remote support activation/deactivation screen
 
Bug Fixes & Changes
- Ethanol Lid pickup failure from Carrier 3 now displays the correct message to users
 - Improved messaging for incorrect number of barcodes during a Gantry Scan
 - Improved audit reporting display
 - Improved device error handling processing and messaging
 - Improved setup information import to deny duplicate sample index entries
 
Previous versions